Armenta_2008_Appl.Spectrosc_62_1322

Reference

Title : A mid-infrared flow-through sensor for label-free monitoring of enzyme inhibition - Armenta_2008_Appl.Spectrosc_62_1322
Author(s) : Armenta S , Tomischko W , Lendl B
Ref : Appl Spectrosc , 62 :1322 , 2008
Abstract :

Label-free monitoring of acetylcholinesterase (AChE) activity was achieved with a mid-infrared flow-through sensor. The flow-through sensor comprised agarose beads, carrying covalently immobilized AChE, which were placed in a temperature-controlled (37 degrees C) CaF(2) flow cell with an optical path of 60 mum. The sensor was incorporated into a computer-controlled sequential injection (SI) system for automated liquid handling. Different mixtures of enzyme substrate acetylcholine (ACh) and inhibitor (tacrine) were prepared and fed into the flow-through sensor. The flow was stopped as soon as the prepared mixtures reached the sensor. Enzymatic hydrolysis of ACh by AChE was directly monitored as it took place in the flow-through sensor. The inhibition effect of tacrine was calculated from the reaction-induced spectral changes, revealing an important decrease in the activity of AChE, approaching zero when the inhibitor concentration is high enough. The developed mid-infrared flow-through sensor is flexible and can be used to study the inhibitor activity of different target molecules as well as different enzymes.

PubMedSearch : Armenta_2008_Appl.Spectrosc_62_1322
PubMedID: 19094390

Related information

Citations formats

Armenta S, Tomischko W, Lendl B (2008)
A mid-infrared flow-through sensor for label-free monitoring of enzyme inhibition
Appl Spectrosc 62 :1322

Armenta S, Tomischko W, Lendl B (2008)
Appl Spectrosc 62 :1322