Title : Methods for detection of anatoxin-a(s) by liquid chromatography coupled to electrospray ionization-tandem mass spectrometry - Dorr_2010_Toxicon_55_92 |
Author(s) : Dorr FA , Rodriguez V , Molica R , Henriksen P , Krock B , Pinto E |
Ref : Toxicon , 55 :92 , 2010 |
Abstract :
Anatoxin-a(s) is a potent irreversible inhibitor of the enzyme acetylcholinesterase with a unique N-hydroxyguanidine methylphosphate ester chemical structure. Determination of this toxin in environmental samples is hampered by the lack of specific methods for its detection. Using the toxic strain of Anabaena lemmermani PH-160 B as positive control, the fragmentation characteristics of anatoxin-a(s) under collision-induced dissociation conditions have been investigated and new LC-MS/MS methods proposed. Recommended ion transitions for correct detection of this toxin are 253>58, 253>159, 235>98 and 235>96. Chromatographic separation is better achieved under HILIC conditions employing a ZIC-HILIC column. This method was used to confirm for the first time the production of anatoxin-a(s) by strains of Anabaena oumiana ITEP-025 and ITEP-026. Considering no standard solutions are commercially available, our results will be of significant use for the correct identification of this toxin by LC-MS/MS. |
PubMedSearch : Dorr_2010_Toxicon_55_92 |
PubMedID: 19632258 |
Inhibitor | Guanitoxin |
Dorr FA, Rodriguez V, Molica R, Henriksen P, Krock B, Pinto E (2010)
Methods for detection of anatoxin-a(s) by liquid chromatography coupled to electrospray ionization-tandem mass spectrometry
Toxicon
55 :92
Dorr FA, Rodriguez V, Molica R, Henriksen P, Krock B, Pinto E (2010)
Toxicon
55 :92