Title : High-performance liquid chromatographic determination of urinary 2,5-hexanedione as mono-2,4-dinitrophenylhydrazone using ultraviolet detection - Gori_1995_J.Chromatogr.B:.Biomedical.Applications_673_165
Author(s) : Gori G , Bartolucci GB , Sturaro A , Parvoli G , Doretti L , Troiano R , Casetta B
Ref : Journal of Chromatography B: Biomedical Applications , 673 :165 , 1995
Abstract :

The good correlation between exposure to n-hexane and 2,5-hexanedione urinary excretion confers on this diketone an important toxicological meaning. this paper proposes a reversed-phase HPLC method which includes, after acid hydrolysis, a derivatization step of 2,5-hexanedione with 2,4-dinitrophenylhydrazine at 70 degrees C for 20 min. The reaction conditions, such as temperature, reagent concentration and time, are optimized so as to allow the condensation of a single carbonyl group. A linear response was obtained in the 0.19-20.0 mg/l range with a detection limit of 0.03 mg/l, corresponding to a signal-to-noise ratio of 3. A phosphate buffer (pH 3.3)-acetonitrile mixture (50:50) as the eluent and UV detection at 334 nm were used.

PubMedSearch : Gori_1995_J.Chromatogr.B:.Biomedical.Applications_673_165
PubMedID: 8611949

Related information

Citations formats

Gori G, Bartolucci GB, Sturaro A, Parvoli G, Doretti L, Troiano R, Casetta B (1995)
High-performance liquid chromatographic determination of urinary 2,5-hexanedione as mono-2,4-dinitrophenylhydrazone using ultraviolet detection
Journal of Chromatography B: Biomedical Applications 673 :165

Gori G, Bartolucci GB, Sturaro A, Parvoli G, Doretti L, Troiano R, Casetta B (1995)
Journal of Chromatography B: Biomedical Applications 673 :165