Title : A disposable amperometric biosensor for rapid screening of anticholinesterase activity in soil extracts - Guerrieri_2002_Analyst_127_5 |
Author(s) : Guerrieri A , Monaci L , Quinto M , Palmisano F |
Ref : Analyst , 127 :5 , 2002 |
Abstract :
A disposable amperometric biosensor for the determination of anticholinesterase activity in soil extracts is described. The sensitive membrane was obtained by co-crosslinking acetylcholinesterase and choline oxidase with bovine serum albumin using glutaraldehyde. The anticholinesterase activity of the soil extract was measured using chronoamperometry at 650 mV vs. Ag/AgCl to monitor the biocatalytically produced H2O2 before and after the inhibition step. An inhibition percentage of 38 +/- 4% was recorded for soil extracts spiked with 10 ppb of ethyl parathion. The device has the potential to be used as a gross sensor for the assessment of anticholinesterase activity in soil extracts. |
PubMedSearch : Guerrieri_2002_Analyst_127_5 |
PubMedID: 11827395 |
Guerrieri A, Monaci L, Quinto M, Palmisano F (2002)
A disposable amperometric biosensor for rapid screening of anticholinesterase activity in soil extracts
Analyst
127 :5
Guerrieri A, Monaci L, Quinto M, Palmisano F (2002)
Analyst
127 :5