Title : Cholinesterase measurements with an automated kit - Oliveira_2002_Am.J.Ind.Med_Suppl_49
Author(s) : Oliveira GH , Henderson JD , Wilson BW
Ref : American Journal of Industrial Medicine , Suppl :49 , 2002
Abstract :

BACKGROUND The Test-mate kit determines acetylcholinesterase (AChE, EC 3.1.1.7) and hemoglobin content of a drop of blood, displaying enzyme activities normalized to 25 degrees C. Previous models produced inconsistent results at different temperatures. This report focuses on the current model, ChE 400, and two instruments of a previous OP model. METHODS: AChE activities were determined by the Ellman assay, using the three kits and a 96-well microplate reader. Temperatures ranged from 10 to 37 degrees C. Fetal bovine serum was the source of AChE. RESULTS: Normalized activities decreased below 20 degrees C in the ChE model and below 25 degrees C in the OP models. Activities of the same serum sample differed between the three Test-mate kits, ranging from 1.03 to 1.49 micromoles/min/ml. Percent errors were greater than with the microplate reader at all temperatures.
CONCLUSIONS: Neither we nor the manufacturer recommend the current Test-mate model for fieldwork. Nevertheless, there have been field measurements with Test-Mate kits, and we recommend that an enzyme activity standard be run in parallel with their use.

PubMedSearch : Oliveira_2002_Am.J.Ind.Med_Suppl_49
PubMedID: 12210682

Related information

Citations formats

Oliveira GH, Henderson JD, Wilson BW (2002)
Cholinesterase measurements with an automated kit
American Journal of Industrial Medicine Suppl :49

Oliveira GH, Henderson JD, Wilson BW (2002)
American Journal of Industrial Medicine Suppl :49