| Title : Measurement of Acetylcholine in Rat Brain Microdialysates by LC - Isotope Dilution Tandem MS - Prokai_2008_Chromatographia_68_s101 |
| Author(s) : Prokai L , Frycak P , Stevens SM, Jr. , Nguyen V |
| Ref : Chromatographia , 68 :s101 , 2008 |
|
Abstract :
An LC-MS/MS method was developed for measuring acetylcholine (ACh) in an aqueous medium using reversed-phase ion-pair chromatography, electrospray ionization on a quadrupole ion trap instrument and a tetradeuterated analogue (ACh-1,1,2,2-d(4)) as an internal standard. A rapid separation was achieved on a 5-cm long octadecylsilica column (2.1 mm i.d.) by employing heptafluorobutyric acid (0.1% v/v) as an ion-pairing agent and requiring 10% v/v acetonitrile in 20 mM ammonium formate buffer under isocratic elution at 200 mul/min flow rate. The instrument's response was calibrated with samples containing known mole ratios of ACh and ACh-1,1,2,2-d(4) in an artificial cerebrospinal fluid, which afforded the conclusion that analyte concentrations could be determined by multiplying the measured analyte to internal standard ion-current ratio with the known molar concentration of the ACh-1,1,2,2-d(4) added. The rapid and simple assay was tested by measuring the basal neurotransmitter concentration in rat brain microdialysates without the use of a cholinesterase inhibitor upon sample collection. |