| Title : Capillary ion electrophoresis screening of nerve agent degradation products in environmental samples using conductivity detection - Rosso_1998_J.Chromatogr.A_824_125 |
| Author(s) : Rosso TE , Bossle PC |
| Ref : Journal of Chromatography A , 824 :125 , 1998 |
|
Abstract :
A method of detecting signature methylphosphonic acid (MPA) breakdown products of V and G nerve agents in environmental samples was developed using capillary ion electrophoresis with conductivity detection. The electrolyte (30 mM L-histidine, 30 mM 2-(N-morpholino)ethanesulfonic acid, 0.7 mM tetradecyltrimethylammonium hydroxide, and 0.03 weight% Triton X-100) allowed baseline separation of MPA, ethyl methylphosphonic acid (EMPA), isopropyl methylphosphonic acid (IMPA), and pinacolyl methylphosphonic acid (PMPA) in less than 10 min. Detector response was linear in the 6-60 micrograms/ml concentration range (correlation coefficient = 0.99) with a detection limit around 6 micrograms/ml. The application of this method for screening MPA, EMPA, IMPA, and PMPA in surface water, groundwater, and soil extracts is demonstrated. |
| PubMedSearch : Rosso_1998_J.Chromatogr.A_824_125 |
| PubMedID: 9818432 |
| Inhibitor | EMPA |
Rosso TE, Bossle PC (1998)
Capillary ion electrophoresis screening of nerve agent degradation products in environmental samples using conductivity detection
Journal of Chromatography A
824 :125
Rosso TE, Bossle PC (1998)
Journal of Chromatography A
824 :125