Stepurska_2015_Anal.Chim.Acta_854_161

Reference

Title : Feasibility of application of conductometric biosensor based on acetylcholinesterase for the inhibitory analysis of toxic compounds of different nature - Stepurska_2015_Anal.Chim.Acta_854_161
Author(s) : Stepurska KV , Soldatkin CO , Kucherenko IS , Arkhypova VM , Dzyadevych SV , Soldatkin AP
Ref : Anal Chim Acta , 854 :161 , 2015
Abstract :

This study was aimed at the development of a conductometric biosensor based on acetylcholinesterase considering the feasibility of its application for the inhibitory analysis of various toxicants. In this paper, the optimum conditions for enzyme immobilization on the transducer surface are selected as well as the optimum concentration of substrate for inhibitory analysis. Sensitivity of the developed biosensor to different classes of toxic compounds (organophosphorus pesticides, heavy metal ions, surfactants, aflatoxin, glycoalkaloids) was tested. It is shown that the developed biosensor can be successfully used for the analysis of pesticides and mycotoxins, as well as for determination of total toxicity of the samples. A new method of biosensor analysis of toxic substances of different classes in complex multicomponent aqueous samples is proposed.

PubMedSearch : Stepurska_2015_Anal.Chim.Acta_854_161
PubMedID: 25479880

Related information

Citations formats

Stepurska KV, Soldatkin CO, Kucherenko IS, Arkhypova VM, Dzyadevych SV, Soldatkin AP (2015)
Feasibility of application of conductometric biosensor based on acetylcholinesterase for the inhibitory analysis of toxic compounds of different nature
Anal Chim Acta 854 :161

Stepurska KV, Soldatkin CO, Kucherenko IS, Arkhypova VM, Dzyadevych SV, Soldatkin AP (2015)
Anal Chim Acta 854 :161