| Title : A solid-phase immobilized epitope immunoassay (SPIE-IA) permitting very sensitive and specific measurement of angiotensin II in plasma - Volland_1999_J.Immunol.Methods_228_37 |
| Author(s) : Volland H , Pradelles P , Ronco P , Azizi M , Simon D , Creminon C , Grassi J |
| Ref : Journal of Immunological Methods , 228 :37 , 1999 |
|
Abstract :
We have developed a new enzyme immunometric assay for angiotensin II (AII) based on SPIE-IA technology (solid-phase immobilized epitope-immunoassay). A monoclonal antibody with optimal properties (mAb3 131) was selected from a series of 19 anti-AII mAbs. The mAb had to be purified from ascitic fluid in a specific manner in order to remove endogenous AII from the antibody-binding sites. We established a sensitive (minimum detectable concentration 0.5 pg/ml) and precise (CV below 15% in the 2-100 pg/ml range) SPIE-IA. Using different AII-related peptides, we observed that this new assay has a specificity profile that compares favourably with the corresponding competitive immunoassay. We have used the assay to measure AII in 42 plasma samples, and demonstrated a good correlation with values obtained using a commercial radioimmunoassay. Assay specificity was supported by HPLC fractionation experiments, confirming the absence of interference induced by endogenous AII-related products. |
| PubMedSearch : Volland_1999_J.Immunol.Methods_228_37 |
| PubMedID: 10556541 |
Volland H, Pradelles P, Ronco P, Azizi M, Simon D, Creminon C, Grassi J (1999)
A solid-phase immobilized epitope immunoassay (SPIE-IA) permitting very sensitive and specific measurement of angiotensin II in plasma
Journal of Immunological Methods
228 :37
Volland H, Pradelles P, Ronco P, Azizi M, Simon D, Creminon C, Grassi J (1999)
Journal of Immunological Methods
228 :37