Wang_2015_Anal.Sci_31_591

Reference

Title : Microfluidic Device for Coulometric Detection of Organophosphate Pesticides - Wang_2015_Anal.Sci_31_591
Author(s) : Wang J , Satake T , Suzuki H
Ref : Anal Sci , 31 :591 , 2015
Abstract :

A microdevice for coulometric detection of organophosphate pesticides (OPs) was fabricated based on the measurement of the inhibition of an enzyme, acetylcholinesterase (AChE), by OPs. Thiocholine (TCh) produced in the enzymatic reaction of AChE with acetylthiocholine (ATCh) as a substrate was oxidized on a microelectrode array formed in a main flow channel. Volumes of plugs of necessary solutions were measured using a structure consisting of a row of rhombuses formed in an auxiliary flow channel. The plugs were merged and solution components were mixed at a T-junction formed with the main and auxiliary flow channels. A linear relationship was confirmed between the generated charge and the logarithm of the OP (malathion) concentration in a concentration range between 10(-6) and 10(-3) M with a correlation coefficient of 0.951. The lower limit of detection was 412 nM.

PubMedSearch : Wang_2015_Anal.Sci_31_591
PubMedID: 26165279

Related information

Citations formats

Wang J, Satake T, Suzuki H (2015)
Microfluidic Device for Coulometric Detection of Organophosphate Pesticides
Anal Sci 31 :591

Wang J, Satake T, Suzuki H (2015)
Anal Sci 31 :591