Title : Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman s Photometric Method - Pohanka_2008_Sensors.(Basel)_8_5229
Author(s) : Pohanka M , Hrabinova M , Kuca K
Ref : Sensors (Basel) , 8 :5229 , 2008
Abstract :

An electrochemical sensor is introduced as a tool applicable for diagnosis of intoxication by cholinesterase inhibitors caused by the well-known nerve agent VX. The traditional Ellman method was chosen for comparison with the sensor's analytical parameters. Both methods are based on estimation of blood cholinesterase inhibition as a marker of intoxication. While Ellman s method provided a limit of detection of 5.2 10-7 M for blood containing VX, the electrochemical sensor was able to detect 4.0 10-7 M. Good correlation between both methods was observed (R = 0.92). The electrochemical sensor could be considered a convenient tool for a fast yet accurate method, easily available for field as well as laboratory use. Time and cost savings are key features of the sensor-based assay.