Pohanka_2007_Acta.Medica.(Hradec.Kralove)_50_239

Reference

Title : Amperometric biosensor for pesticide methamidophos assay - Pohanka_2007_Acta.Medica.(Hradec.Kralove)_50_239
Author(s) : Pohanka M , Kuca K , Jun D
Ref : Acta Medica (Hradec Kralove) , 50 :239 , 2007
Abstract :

Amperometric biosensor based on enzyme acetylcholinesterase (AChE; EC 3.1.1.7) was tested for pesticide methamidophos assay. Biosensor consists from four screen printed platinum electrodes on ceramic strip. AChE was physically adsorbed onto the electrode surface. The measuring principle was based on the inhibition of AChE activity in the presence of methamidophos. The proposed method limit of detection was 2.45 nM, responding to 3.46 pg of methamidophos detected absolutely when we consider the sample volume.

PubMedSearch : Pohanka_2007_Acta.Medica.(Hradec.Kralove)_50_239
PubMedID: 18290547

Related information

Citations formats

Pohanka M, Kuca K, Jun D (2007)
Amperometric biosensor for pesticide methamidophos assay
Acta Medica (Hradec Kralove) 50 :239

Pohanka M, Kuca K, Jun D (2007)
Acta Medica (Hradec Kralove) 50 :239